Beckman Instrumentation Laboratory

ACL 10000 Coagulation Analyzer
Model : ACL 10000
LINEARITY RANGE APTT, u hep/mL blood (PLASMA SAMPLE) |
Reagent dependent |
FIB, mg/dl (PLASMA SAMPLE) |
40-800, 30-1100 for Clauss Fibrinogen |
POWER NEEDED, VAC (ENERGY) |
110/220 |
Type (CONFIGURATION) |
Benchtop |
PRINTER (EXTERNAL BEAM) |
Yes, internal or external – for details please see comments80 columns, text interface (external) |
DIMENSIONS (HXWXD) CM, (IN) (DISPLAY) |
60 x 92 x 60 (23.6 x 36.2 x 23.6) |
WEIGHT, kg (lb) (DISPLAY) |
63 (139) |
STAT CAPABILITY (PLASMA SAMPLE) |
Yes |
CENTRIFUGE NEEDED (APTT INCUBATION) |
Yes |
PRIMARY TUBE SAMPLING (APTT INCUBATION) |
Yes |
COMPUTER INTERFACE |
RS232C |
DISPLAY |
LCD – for more information please see commentstouchscreen |
BAR CODE |
Yes |
PLASMA SAMPLE |
|
Max sample onboard (PLASMA SAMPLE) |
40 |
Size, ml (PLASMA SAMPLE) |
Variable, <0.1 |
YIELD, RESULTS/HR (APTT INCUBATION) |
175 PTs w/ratios and 175 FIBs |
MANUFACTURER REAGENTS (APTT INCUBATION) |
Yes |
APTT INCUBATION |
|
Sample time flag (APTT INCUBATION) |
Yes |
Time, min (APTT INCUBATION) |
5 |
AUTO SAMPLE HANDLING (PLASMA SAMPLE) |
Yes |
DETECTION TECHNIQUE (APTT INCUBATION) |
Nephelometric, photo-optical |
OTHER ATTRIBUTES (Interference compensation) |
Database onboard; QC onboard; automatic dilutions; stored calibration curves; alphanumeric keypad; profiles of routine tests; optional Windows Research PLUS software; reflex testing. Approved by CSA and TUV. |
FDA CLEARANCE (Interference compensation) |
Yes |
CE MARK (MDD) (Interference compensation) |
Yes |
MARKETING REGION (Interference compensation) |
Worldwide |
Full Automation |
Yes |
Weight: 139 Lbs.
Dimensions: 24(H) x 36(W) x 24(D) inches

ACL 7000 Coagulation Analyzer
Model : ACL 7000
Beckman/Instrumentation Laboratory ACL 7000 Coagulation Analyzer Highlights
· Performs clotting, chromogenic and immunological testing
· Accepts primary tubes
· Includes integrated sample barcode reader
· Has onboard database for patient, QC and calibration results
· Allows for random patient programming via bi-directional host interface
Ideal for lower-volume hemostasis labs with basic hemostasis testing needs, yet offers the same high quality, reliability and accuracy as larger ACL models.
Weight: 114 Lbs.
Dimensions: 18(H) x 30(W) x 27(D) inches

ACL 8000 Coagulation Analyzer
Model : ACL 8000
Beckman/Instrumentation Laboratory ACL 8000 Coagulation Analyzer Highlights
LINEARITY RANGE APTT, u hep/mL blood (PLASMA SAMPLE) |
Reagent dependent |
FIB, mg/dl (PLASMA SAMPLE) |
40-800, 30-1100 for Clauss Fibrinogen |
POWER NEEDED, VAC (ENERGY) |
110/220 |
Type (CONFIGURATION) |
Benchtop |
DISPLAY |
LCD – for more information please see commentstouchscreen |
YIELD, RESULTS/HR (APTT INCUBATION) |
175 PTs w/ratios and 175 FIBs |
MANUFACTURER REAGENTS (APTT INCUBATION) |
Yes |
CENTRIFUGE NEEDED (APTT INCUBATION) |
Yes |
STAT CAPABILITY (PLASMA SAMPLE) |
Yes |
APTT INCUBATION |
|
Sample time flag (APTT INCUBATION) |
Yes |
Time, min (APTT INCUBATION) |
5 |
COMPUTER INTERFACE |
RS232C |
BAR CODE |
Yes |
AUTO SAMPLE HANDLING (PLASMA SAMPLE) |
Yes |
PLASMA SAMPLE |
|
Max sample onboard (PLASMA SAMPLE) |
40 |
Size, ml (PLASMA SAMPLE) |
Variable, <0.1 |
PRIMARY TUBE SAMPLING (APTT INCUBATION) |
Yes |
PRINTER (EXTERNAL BEAM) |
Yes, internal or external – for details please see comments80 columns, text interface (external) |
DIMENSIONS (HXWXD) CM, (IN) (DISPLAY) |
60 x 92 x 60 (23.6 x 36.2 x 23.6) |
WEIGHT, kg (lb) (DISPLAY) |
63 (139) |
DETECTION TECHNIQUE (APTT INCUBATION) |
Nephelometric, photo-optical |
OTHER ATTRIBUTES (Interference compensation) |
Database onboard; QC onboard; library of 300 tests; automatic dilutions; stored calibration curves; reaction curves; touchscreen stat capability; reflex testing. Approved by CSA and TUV. |
FDA CLEARANCE (Interference compensation) |
Yes |
CE MARK (MDD) (Interference compensation) |
Yes |
MARKETING REGION (Interference compensation) |
Worldwide |
Full Automation |
Yes |
Weight: 139 Lbs.
Dimensions: 24(H) x 36(W) x 24(D) inches

ACL 9000 Coagulation Analyzer
Model : ACL 9000
Beckman/Instrumentation Laboratory ACL 9000 Coagulation Analyzer Highlights
· Extensive test menu with clotting, chromogenic and immunologic capabilities
· Clot curve analysis evaluates abnormal or unexpected test results
· Reflex testing standardizes the decision rules
· Automated rotor changer delivers 18 tests in 7.5 minutes
· Easy-to-use touch screen
· Continuous loading with STAT interrupt
· On board capacity for 40 samples and 18 reagents
Sample Area
The removable sample tray holds up to 40 patient samples for superior walk-away capability. This system allows you to use primary tubes of different vial dimensions and cups. Plus, the on-board bar-code reader takes care of the sample identification. The continuous loading feature maximizes access to the analyzer, allowing you to add more samples whenever you are ready.
Reagent Area
Up to 18 reagent vials may be placed in two locations on the ACL 9000. There are 8 positions which hold original reagent vials of different sizes, from the largest 16 mL fill to the smallest 4 mL fill. Four of the positions are for reagents that require refrigeration and stirring and other 4 positions are kept at ambient temperature. A second reagent area includes 10 additional positions that accept either 10 mL vials and cups.
The ACL 9000 can monitor in real time the amount of reagent in the vials using a liquid level detection system and the reagent configuration can be saved by designing “reagent maps” for an unlimited number of test combinations.
Cuvette Area
ACL 9000 provides superior system autonomy with on-board capacity of 260 reaction cuvettes –13 rotors of 20 cuvettes each. The rotors, which may be added at the user’s convenience, are automatically transported to and from the reaction measuring area by means of a slider and a robotic arm mechanism. This transfers unused rotors from their temperature controlled storage area to the area where loading and analysis take place and, once analysis is completed, from the analytical area to the waste rotor container. The rotor waste system is operator safe allowing easy removal of used rotors without interruption of system operation.
Wash-Reference Emulsion System
The larger 1-liter bottle of Wash-Reference Emulsion increases the system autonomy. A sensor that provides real time information to the user monitors the liquid level. This proven dilutor technology used in the ACL family of systems provides a high level of precision and accuracy for sample and reagent dispensation.
Weight: 138 Lbs.
Dimensions: 11(H) x 36(W) x 24(D) inches

ACL Elite Coagulation Analyzer
Model : ACL Elite
Beckman/Instrumentation Laboratory ACL Elite Coagulation Analyzer Highlights
Benefits of the ACL ELITE PRO include:
· Full routine test menu, including automated D-Dimer and Heparin anti-Xa
· Rapid assay turnaround times
· Easy-to-use “Load & Run” operation
· Intuitive, icon-based software that helps maximize training time
· True walk-away capability allowing any time STAT access
· Non-interrupted loading of reagents and consumables to maximize productivity
· Bar-coded sample and reagent capability for efficient processing
· Require less than 10 minutes of daily, weekly and monthly maintenance
Weight: 139 Lbs.
Dimensions:
Height: 24 inches
Width: 36 inches
Depth: 24 inches

ACL Top 300, 500 and 700 Coagulation Analyzer
Model : ACL Top 300, 500 and 700
Beckman/Instrumentation Laboratory ACL Top 300, 500 and 700 Coagulation Analyzer Highlights
User-Friendly System for Increased Efficiency
· Easy to operate with complete walkaway capabilities for improved productivity all day, every day
· Continuous loading and unloading of samples, reagents and cuvettes without interruption
· Simultaneously analyzes specimens, monitors reagents and performs hardware diagnostics
· Intuitive operating system with user-friendly software and easy to use touch screen monitor
· Sophisticated software scheduling optimizes testing
· Bar-coded reagents allow for fast, simple loading; scanner reads multiple bar-codes simultaneously
· On-board operator’s manual and help screen provide information quickly and conveniently
· Minimal maintenance allows for increased uptime
· Modern ergonomic design for a safe, functional work environment
Advanced Features for Improved Patient Care
· STAT samples may be introduced at any time, in any rack and in any position without disrupting operation
· Rerun and Reflex testing enable labs to automatically process abnormal or unexpected sample results
· Clot signature curves help troubleshoot abnormal samples and gather additional clinical information to aid in patient assessment
· Factor Parallelism aids in the analysis of factor assays while helping the laboratory determine the presence of inhibitors
· On-board Quality Control program ensures reliable results
Weight: 344 Lbs.
Dimensions:
Height 59 inches
Weight 30 inches
Depth 29 inches